4.7 Article

Liver iron concentration quantification by MRI: are recommended protocols accurate enough for clinical practice?

Journal

EUROPEAN RADIOLOGY
Volume 21, Issue 1, Pages 137-141

Publisher

SPRINGER
DOI: 10.1007/s00330-010-1899-z

Keywords

Haemochromatosis; Liver biopsy; Liver diseases; Magnetic resonance imaging; Comparative studies

Ask authors/readers for more resources

To assess the accuracy of quantification of liver iron concentration (LIC) by MRI using the Rennes University (URennes) algorithm. In the overall study period 1999-2006 the LIC in 171 patients was calculated with the URennes model and the results were compared with LIC measured by liver biopsy. The biopsy showed that 107 patients had no overload, 38 moderate overload and 26 high overload. The correlation between MRI and biopsy was r = 0.86. MRI correctly classified 105 patients according to the various levels of LIC. Diagnostic accuracy was 61.4%, with a tendency to overestimate overload: 43% of patients with no overload were diagnosed as having overload, and 44.7% of patients with moderate overload were diagnosed as having high overload. The sensitivity of the URennes method for high overload was 92.3%, and the specificity for the absence of overload was 57.0%. MRI values greater than 170 mu mol Fe/g revealed a positive predictive value (PPV) for haemochromatosis of 100% (n = 18); concentrations below 60 mu mol Fe/g had a negative predictive value (NPV) of 100% for haemochromatosis (n = 101). The diagnosis in 44 patients with intermediate values remained uncertain. The assessment of LIC with the URennes method was useful in 74.3% of the patients to rule out or to diagnose high iron overload. The method has a tendency to overestimate overload, which limits its diagnostic performance.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available