Journal
ELECTROPHORESIS
Volume 32, Issue 5, Pages 494-505Publisher
WILEY
DOI: 10.1002/elps.201000390
Keywords
Dielectrophoretic manipulation; Electric field gradient distribution; MEMS; Optimal design; Three-dimensional microelectrode
Funding
- National Science Foundation of China [50875103, 90923019]
- Foundation from Wuhan National Laboratory for Optoelectronics
Ask authors/readers for more resources
Dielectrophoresis (DEP) has been deemed as a potential and ideal solution for bioparticle manipulation. A 3-D carbon micro-electro-mechanical system (MEMS) fabricated from the latest developed carbon-MEMS approach has advantages of offering low-cost, biocompatible and high-throughput DEP manipulation for bioparticles. In this paper, a typical process for fabrication of various 3-D microelectrode configurations was demonstrated; accurate numerical analysis was presented on electric field gradient distribution and DEP force based on various microelectrode array configurations. The effects of electrode edge angle, electrode edge-to-edge spacing and electrode height on the electric field distributions were investigated, and optimal design considerations and rules were concluded through analysis of results. The outcomes demonstrate that the sharp edge electrode is more effective in DEP manipulation and both electrode edge-to-edge spacing and electrode height are critical design parameters for seeking optimal DEP manipulation. The gradient magnitude increases exponentially as the electrode spacing is reduced and the electric field extends significantly as the electrode height increases, both of which contribute to a higher throughput for DEP manipulation. These findings are consistent with experimental observations in the literature and will provide critical guidelines for optimal design of DEP devices with 3-D carbon-MEMS.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available