4.3 Article

BioFET sensor for detection of albumin in urine

Journal

ELECTRONICS LETTERS
Volume 44, Issue 3, Pages 185-U7

Publisher

INST ENGINEERING TECHNOLOGY-IET
DOI: 10.1049/el:20083551

Keywords

-

Ask authors/readers for more resources

Characterisation of a BioFET for detection of albumin in a mixture of human urine is presented. To avoid electrolyte effect of the urine, it was measured in PBS (phosphate buffer saline) at a fixed pH after albumin binding. The drain current was modulated by the albumin bound to the anti-albumin immobilised on the gate surface of the BioFET. The current variation ratio was likely to be proportional to the concentration of the albumin in the range 50 - 250 mg/1. The results show the feasibility of the BioFET as a urinary albumin sensor.