4.6 Article Proceedings Paper

Effect of oxygen plasma treatment on non-thrombogenicity of diamond-like carbon films

Journal

DIAMOND AND RELATED MATERIALS
Volume 19, Issue 7-9, Pages 861-865

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.diamond.2010.02.003

Keywords

Diamond-like carbon; Plasma CVD; Surface characterization; Biomedical applications

Ask authors/readers for more resources

The non-thrombogenicity of oxygen-plasma-treated DLC films was investigated as surface coatings for medical devices. DLC films were deposited on polycarbonate substrates by a radio frequency plasma enhanced chemical vapor deposition method using acetylene gas. The deposited DLC films were then treated with plasma of oxygen gas at powers of 15W, 50W, and 200W. Wettability was evaluated by water contact angle measurements and the changes in surface chemistry and roughness were examined by X-ray photoelectron spectroscopy and atomic force microscope analysis, respectively. Each oxygen-plasma-treated DLC film exhibited a hydrophilic nature with water contact angles of 11.1 degrees, 17.7 degrees and 36.8 degrees. The non-thrombogenicity of the samples was evaluated through the incubation with platelet-rich plasma isolated from human whole blood. Non-thrombogenic properties dramatically improved for both 15 W- and 50 W-oxygen-plasma-treated DLC films. These results demonstrate that the oxygen plasma treatment at lower powers promotes the non-thrombogenicity of DLC films with highly hydrophilic surfaces. (C) 2010 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available