4.2 Article

Nonlinear Stress Analysis of Titanium Implants by Finite Element Method

Journal

DENTAL MATERIALS JOURNAL
Volume 27, Issue 4, Pages 633-639

Publisher

JAPANESE SOC DENTAL MATERIALS DEVICES
DOI: 10.4012/dmj.27.633

Keywords

Finite element method; Titanium implant; Nonlinear stress analysis

Ask authors/readers for more resources

With use of dental implants on the rise, there is also a tandem increase in the number of implant fracture reports. To the end of investigating the stress occurring in implants, elasticity and plasticity analyses were performed using the finite element method. The following results were obtained: (1) With one-piece type of implants of 3.3 mm diameter, elasticity analysis showed that after applying 500 N in a 45-degree direction. stress exceeding 500 MPa - which is the proof stress of grade 4 pure titanium - occured. This suggested the possibility of fatigue destruction due to abnormal occlusal force, such as during bruxism. (2) With two-piece type of implants that can tolerate vertical loading of 5,000 N, plasticity analysis suggested the possibility of screw area fracture after applying 500 N in a 45-degree direction. (3) On the combined use of all abutment, and a fixture from different manufacturers, fracture destruction of even Ti-6Al-4V. which has a high degree of strength, was predicted.