4.5 Article

Construction and calibration of a low cost and fully automated vibrating sample magnetometer

Journal

JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS
Volume 386, Issue -, Pages 25-30

Publisher

ELSEVIER
DOI: 10.1016/j.jmmm.2015.03.051

Keywords

Vibrating sample magnetometer; Magnetic materials; Magnetic hysteresis; Data acquisition

Ask authors/readers for more resources

A low cost vibrating sample magnetometer (VSM) has been constructed by using an electromagnet and an audio loud speaker; where both are controlled by a data acquisition device. The constructed VSM records the magnetic hysteresis loop up to 8.3 KG at room temperature. The apparatus has been calibrated and tested by using magnetic hysteresis data or some ferrite samples measured by two scientifically calibrated magnetometers; model (Lake Shore 7410) and model (LDJ Electronics Inc. Troy, MI). Our VSM lab-built new design proved success and reliability. (C) 2015 Published by Elsevier B.V.