Performance evaluation of a particle-enhanced turbidimetric cystatin C assay on the Abbott ci8200 analyzer

Journal

CLINICAL BIOCHEMISTRY
Volume 42, Issue 9, Pages 873-876

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2009.02.024

Keywords

Cystatin C; Glomerular filtration rate; Human; Kidney

Funding

  1. Uppsala University Hospital Research Fund

Ask authors/readers for more resources

Objectives: Glomerular filtration rate (GFR) is widely accepted as the best overall measure of kidney function. Cystatin C is a novel endogenous GFR marker that has been shown to be superior to creatinine for estimation of GFR in several studies. There is a need for cystatin C assays adapted to routine chemistry instrument to minimize turnaround times and allowing 24 h/day availability. Materials and methods: We have evaluated a new cystatin C assay developed for Architect cSystem (Abbott Laboratories, Abbott Park, IL, USA). Results: The cystatin C assay showed good agreement with the corresponding assay from Dade Behring (Deerfield, IL, USA). The assay has a very low total imprecision and a good linearity. Conclusions: The new cystatin C assay is an interesting alternative to current cystatin C assays. On an Architect cSystem the assay can be performed with the same turnaround times and availability as creatinine. (c) 2009 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available