Effects of Neurally Adjusted Ventilatory Assist (NAVA) levels in non-invasive ventilated patients: titrating NAVA levels with electric diaphragmatic activity and tidal volume matching

Journal

BIOMEDICAL ENGINEERING ONLINE
Volume 12, Issue -, Pages -

Publisher

BIOMED CENTRAL LTD
DOI: 10.1186/1475-925X-12-61

Keywords

Mechanical ventilation; NAVA; Non-invasive ventilation; Patient-ventilator interaction; Matching

Funding

  1. FNRS 474 (Belgium)
  2. FRST (New Zealand)
  3. University of Liege
  4. 475, Belgian French Community (ARC-Academie Wallonie Europe)

Ask authors/readers for more resources

Background: Neurally adjusted ventilatory assist (NAVA) delivers pressure in proportion to diaphragm electrical activity (Eadi). However, each patient responds differently to NAVA levels. This study aims to examine the matching between tidal volume (Vt) and patients' inspiratory demand (Eadi), and to investigate patient-specific response to various NAVA levels in non-invasively ventilated patients. Methods: 12 patients were ventilated non-invasively with NAVA using three different NAVA levels. NAVA100 was set according to the manufacturer's recommendation to have similar peak airway pressure as during pressure support. NAVA level was then adjusted +/- 50% (NAVA50, NAVA150). Airway pressure, flow and Eadi were recorded for 15 minutes at each NAVA level. The matching of Vt and integral of Eadi (.Eadi) were assessed at the different NAVA levels. A metric, Range90, was defined as the 5-95% range of Vt/fEadi ratio to assess matching for each NAVA level. Smaller Range90 values indicated better matching of supply to demand. Results: Patients ventilated at NAVA50 had the lowest Range90 with median 25.6 uVs/ml [Interquartile range (IQR): 15.4-70.4], suggesting that, globally, NAVA50 provided better matching between.Eadi and Vt than NAVA100 and NAVA150. However, on a per-patient basis, 4 patients had the lowest Range90 values in NAVA100, 1 patient at NAVA150 and 7 patients at NAVA50. Robust coefficient of variation for.Eadi and Vt were not different between NAVA levels. Conclusions: The patient-specific matching between fEadi and Vt was variable, indicating that to obtain the best possible matching, NAVA level setting should be patient specific. The Range90 concept presented to evaluate Vt/fEadi is a physiologic metric that could help in individual titration of NAVA level.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available