4.6 Article

Evaluating the Precision of Preoperative Planning in Patient Specific Instrumentation: Can a Single MRI Yield Different Preoperative Plans?

Journal

JOURNAL OF ARTHROPLASTY
Volume 30, Issue 7, Pages 1250-1253

Publisher

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2015.02.021

Keywords

total knee arthroplasty; patient specific instrumentation; component sizing; precision; preoperative planning

Categories

Ask authors/readers for more resources

If PSI preoperative planning were perfectly precise, a single MRI would give rise to one preoperative plan. Our purpose was to determine whether a single MRI inputted into two different PSI software yielded differences in preoperative alignment determination, component sizing, and bone resection within the preoperative plan. This prospective comparative study evaluated 40 preoperative plans generated by two PSI software given identical MRI. Femoral and tibial component sizes differed between software in 37.5% and 30.0% of cases, respectively. The maximum difference in bone resection between software ranged from 2.2 mm to 5.1 mm. Surgeons should be prepared to intraoperatively deviate from PSI selected size by 1 size. It may be necessary to fine tune soft tissue balancing when using a PSI system. (C) 2015 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available