4.7 Article

Evaluation of nanoscale roughness measurements on a plasma treated SU-8 polymer surface by atomic force microscopy

Journal

APPLIED SURFACE SCIENCE
Volume 254, Issue 22, Pages 7290-7295

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.apsusc.2008.05.323

Keywords

AFM; tip degradation; SU-8; plasma treatment; photoresist; hydrophilisation; polymer; intermittent contact

Funding

  1. German Federal Ministry of Education and Research (BMBF) [03N8706]

Ask authors/readers for more resources

A comparison between roughness data obtained with an atomic force microscope (AFM) on different surfaces requires reliable roughness parameters. In order to specify the appropriate parameters for nanoscale roughness measurements, we compared the root mean square (rms) roughness and the relative surface area (sdr) as function of varying scan size, speed and pixel size. By using oxygen plasma (24 kJ) treated SU-8 with an average rms roughness of 2.6 +/- 0.5 nm as reference surface, the repeatability of the method was evaluated for dynamic (tapping) and contact mode. The evaluation of AFM images indicated a decrease of the effective tip radius after a few measurements. This degradation of the tip lowers the resolution of the image and can affect roughness measurements. (C) 2008 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available