Journal
APPLIED PHYSICS LETTERS
Volume 101, Issue 2, Pages -Publisher
AMER INST PHYSICS
DOI: 10.1063/1.4736731
Keywords
-
Categories
Funding
- National Institutes of Health (NIH) [R01-EB12058, P41-EB02182]
Ask authors/readers for more resources
A miniature focused needle transducer (<1 mm) was fabricated using the press-focusing technique. The measured pulse-echo waveform showed the transducer had center frequency of 57.5 MHz with 54% bandwidth and 14 dB insertion loss. To evaluate the performance of this type of transducer, in vitro ultrasonic biomicroscopy imaging on the rabbit eye was obtained. Moreover, a single beam acoustic trapping experiment was performed using this transducer. Trapping of targeted particle size smaller than the ultrasonic wavelength was observed. Potential applications of these devices include minimally invasive measurements of retinal blood flow and single beam acoustic trapping of microparticles. (C) 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4736731]
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available