4.6 Article

An on-line separation and preconcentration system coupled with flame atomic absorption spectrometry for the determination of lead

Journal

ANALYTICAL METHODS
Volume 5, Issue 13, Pages 3307-3313

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3ay40271a

Keywords

-

Ask authors/readers for more resources

A simple on-line separation/preconcentration procedure is described for the determination of trace levels of Pb(II) by flame atomic absorption spectrometry (FAAS). A mini-column packed with Dowex Marathon C (DMC) ion-exchange resin was used for the on-line preconcentration of Pb(II) at pH 3.5. Pb(II) was sorbed on the resin, from which it could be eluted with 3 mol L-1 HCl and then introduced directly to the nebulizer-burner system of FAAS. The parameters influential on the determination of Pb(II) ions such as the pH of sample solution, eluent type, interfering ions and flow variables were studied. Under the optimum conditions, the calibration graph obtained was linear over the concentration range of 0.01-0.1 mg L-1. The detection limit of the method was 1.3 mu g L-1 while precision was 1.1% (n = 15) at a 0.05 mg L-1 Pb(II) level. The accuracy of the method was proven using standard reference materials. The developed method has been applied successfully to the determination of lead in water and various environmental samples with satisfactory results.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings