Ask authors/readers for more resources
A partially preanodized screen-printed carbon electrode (PSPCE*) coupled with flow injection analysis (FIA) was developed to raise the selectivity of ellagic acid (EA). To confirm the effectiveness of partial preanodization, two pretreated screen-printed carbon electrodes were electrochemically compared. One was a PSPCE* fabricated by potential cycling (-1.0-+1.0 V vs. Ag/AgCl) to make the electrode surface partially preanodized and the other was a PSPCE* fabricated by a high treatment potential (+2.0 V vs. Ag/AgCl). Cyclic voltammograms showed that the catalytic current of EA was observed at both the PSPCE* and SPCE*. No catalytic current of ascorbic acid 2-glucoside (AA2G) and magnesium ascorbyl phosphate (MAP) was observed at the PSPCE*. The PSPCE* selectively detected EA. The factors, which influence the EA response current, have previously been discussed. At the detection limit (0.012 ppm, SIN = 3), the linear calibration plot (R-2 = 0.998) was attained for 0.1-50 ppm of EA solutions. A relative standard deviation of 4.37 and 3.90% was conducted for consecutive injections (n = 10) of 1 and 50 ppm EA, respectively. Finally, a practical application of the proposed method was demonstrated by the quantitative analysis of EA in skin whitening creams, and good recovery was obtained.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available