Comparison of Keratometry Obtained by a Swept Source OCT-Based Biometer with a Standard Optical Biometer and Scheimpflug Imaging

Journal

CURRENT EYE RESEARCH
Volume 43, Issue 7, Pages 882-888

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/02713683.2018.1458881

Keywords

Cataract; keratometry; Scheimpflug imaging; swept source OCT-based biometry

Categories

Ask authors/readers for more resources

Purpose: To assess agreement of a swept source-optical coherence tomography (SS-OCT) based Biometer with a standard IOLMaster device and Scheimpflug Imaging (SI) to acquire keratometric measurements in cataract patients. Methods: In this prospective comparative study, 101 eyes of 101 cataract surgery candidates, aged 24-81years, were sequentially examined using three devices. Keratometry values at the flat (K1) and steep (K2) axis, mean corneal power (Km) and magnitude of corneal astigmatism as well as J0 and J45 vectoral components of astigmatism obtained with the SS-OCT based biometer (IOLMaster 700) were compared with those obtained with the IOLMaster 500 and SI. The agreement between measurements was evaluated by the Bland-Altman method, intraclass correlation coefficients (ICCs) and repeated-measures analysis of variance. Results: Mean K1 values from the three devices were similar (p=0.09). Mean K2 and Km values of IOLMaster 700 were higher than SI and lower than IOLMaster 500 (p=0.04 for K2 and p=0.02 for Km). There was a strong correlation between K1, K2, Km and magnitude of astigmatism obtained with all devices (r >= 0.80 and p<0.01). The 95% limits of agreement (LoA) width for each keratometric value were highest for the comparison between IOLMaster 500 and SI and lowest for the comparison between IOLMaster 700 and 500. The mean differences (width of 95% LoA) for J0 and J45 vectoral components were 0.005 (2.19) and 0.12 (2.92) for the measurements obtained by IOLMaster 700 vs IOLMaster 500 and 0.06 (1.79) and 0.02 (1.58) for the measurements obtained by IOLMaster 700 vs SI, respectively. Conclusions: With ICCs close to 1, the agreement between all devices was excellent for keratometric measurements. Mean K2, Km and astigmatism measurements from IOLMaster 700 were lower than IOLMaster 500 and higher than SI. However, the differences were quite small and are not expected to affect the final IOL power.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available