4.6 Review

Radiotherapy in patients with pacemakers and implantable cardioverter defibrillators: a literature review

Journal

EUROPACE
Volume 18, Issue 4, Pages 479-491

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/europace/euv135

Keywords

Pacemaker; Implantable cardioverter defibrillator; Radiotherapy; Device malfunction; Epidemiology; Safety

Ask authors/readers for more resources

An increasing number of patients with implantable cardiac rhythm devices undergo radiotherapy (RT) for cancer and are thereby exposed to the risk of device failure. Current safety recommendations seem to have limitations by not accounting for the risk of pacemakers and implantable cardioverter defibrillators malfunctioning at low radiation doses. Besides scant knowledge about optimal safety measures, only little is known about the exact prevalence of patients with devices undergoing RT. In this review, we provide a short overview of the principles of RT and present the current evidence on the predictors and mechanisms of device malfunctions during RT. We also summarize practical recommendations from recent publications and from the industry. Strongly associated with beam energy of photon RT, device malfunctions occur at similar to 3% of RT courses, posing a substantial issue in clinical practice. Malfunctions described in the literature typically consist of transient software disturbances and only seldom manifest as a permanent damage of the device. Through close cooperation between cardiologists and oncologists, a tailored individualized approach might be necessary in this patient group in waiting time for updated international guidelines in the field.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available