4.5 Article

Ultrafast gas chromatography method with direct injection for the quantitative determination of benzene, toluene, ethylbenzene, and xylenes in commercial gasoline

Journal

JOURNAL OF SEPARATION SCIENCE
Volume 40, Issue 7, Pages 1508-1515

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.201601174

Keywords

direct injection; gas chromatography; gasoline; quality control

Funding

  1. PIBITI/CNPq/IFSP

Ask authors/readers for more resources

Benzene, toluene, ethylbenzene, and xylenes are some of the most hazardous constituents found in commercial gasoline samples; therefore, these components must be monitored to avoid toxicological problems. We propose a new routine method of ultrafast gas chromatography coupled to flame ionization detection for the direct determination of benzene, toluene, ethylbenzene, and xylenes in commercial gasoline. This method is based on external standard calibration to quantify each compound, including the validation step of the study of linearity, detection and quantification limits, precision, and accuracy. The time of analysis was less than 3.2 min, with quantitative statements regarding the separation and quantification of all compounds in commercial gasoline samples. Ultrafast gas chromatography is a promising alternative method to official analytical techniques. Government laboratories could consider using this method for quality control.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available