4.7 Article

Miniaturized Impedance Flow Cytometer: Design Rules and Integrated Readout

Journal

IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS
Volume 11, Issue 6, Pages 1438-1449

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBCAS.2017.2748158

Keywords

Dielectrophoretic cell separation; impedance flow cytometry; microfluidics; resistive pulse detection

Funding

  1. Fondazione Fratelli Agostino and Enrico Rocca

Ask authors/readers for more resources

A dual-channel credit-card-sized impedance cell counter featuring a throughput of 2000 cell/s and detection of single yeast cells (5 mu m) with a signal-to-noise ratio of 20 dB is presented. Its compactness is achieved by a CMOS ASIC combining a lock-in impedance demodulator with an oversampling 20-bit Sigma Delta ADC and real-time peak detection embedded in field-programmable gate array. The module is coupled to a dielectrophoretic cell-sorting microfluidic device, offering compact and label-free electrical readout that replaces the need for a fluorescence microscope and, thus, is suitable for point-of-care diagnostics. The independent role of each dimension of the planar sensing microelectrodes is demonstrated, with simulations and experiments, along with its relevant effect on the spectrum of thin channels, deriving useful design guidelines.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available