The biomechanical effect of anteversion and modular neck offset on stress shielding for short-stem versus conventional long-stem hip implants

Journal

MEDICAL ENGINEERING & PHYSICS
Volume 38, Issue 3, Pages 232-240

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2015.12.005

Keywords

Biomechanics; Stress shielding; Short-stem; Hip arthroplasty; Finite element analysis

Funding

  1. Mr. and Mrs. Frank and Barbara Milligan Graduate Fellowship, the University of Toronto, and St. Michael's Hospital

Ask authors/readers for more resources

Short-stem hip implants are increasingly common since they preserve host bone stock and presumably reduce stress shielding by improving load distribution in the proximal femur. Stress shielding may lead to decreased bone density, implant loosening, and fracture. However, few biomechanical studies have examined short-stem hip implants. The purpose of this study was to compare short-stem vs. standard length stemmed implants for stress shielding effects due to anteversion-retroversion, anterior-posterior position, and modular neck offset. Twelve artificial femurs were implanted with either a short-stem modular-neck implant or a conventional length monolithic implant in 0 degrees or 15 degrees of anteversion. Three modular neck options were tested in the short-stem implants. Three control femurs remained intact. Femurs were mounted in adduction and subjected to axial loading. Strain gauge values were collected to validate a Finite Element (FE) model, which was used to simulate the full range of physiologically possible anteversion and anterior-posterior combinations (n = 25 combinations per implant). Calcar stress was compared between implants and across each implant's range of anteversion using one and two-way ANOVA. Stress shielding was defined as the overall change in stress compared to an intact femur. The FE model compared well with experimental strains (intact: slope = 0.898, R = 0.943; short-stem: slope = 0.731, R = 0.948; standard-stem: slope = 0.743, R = 0.859); correction factors were used to adjust slopes to unity. No implant anteversion showed significant reduction in stress shielding (alpha = 0.05, p > 0.05). Stress shielding was significantly higher in the standard-stem implant (63% change from intact femur, p < 0.001) than in short-stem implants (29-39% change, p < 0.001). Short-stem implants reduce stress shielding compared to standard length stemmed implants, while implant anteversion and anterior-posterior position had no effect. Therefore, short-stem implants have a greater likelihood of maintaining calcar bone strength in the long term. (C) 2015 IPEM. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available