Spatial-resolved electrochemiluminescence ratiometry based on bipolar electrode for bioanalysis

Journal

BIOSENSORS & BIOELECTRONICS
Volume 86, Issue -, Pages 683-689

Publisher

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2016.07.067

Keywords

Electrochemiluminescence; Ratiometry; Resonance energy transfer; Electron transfer; Bipolar electrode; PSA detection

Funding

  1. Ministry of Science and Technology Program of China [2016YFA0201200]
  2. National Natural Science Foundation of China [21327902, 21475058]
  3. Priority Academic Program Development of Jiangsu Higher Education Institutions

Ask authors/readers for more resources

Herein, a, spatial-resolved electrochemiluminescene (ECL) ratiometry based on a closed biopolar electrode (BPE) is reported for the highly sensitive detection of prostate specific antigen (PSA). Au@g-C3N4 NCs as one ECL emitter were firstly coated on the cathode of BPE, while the anode of the BPE served for calibration via another ECL substance, Ru(bpy)(3)(2+). The electroneutrality across the BPE makes the reactions on each pole of BPE electrically coupled. Thus one electrochemical sensing reaction at one pole of BPE could be quantified at both ends. A composite, Pt-PAMAM-DNAzyme was assembled on the surface of cathode via DNA hybridization between probe DNA and PSA aptamer. It acted as an ECL quencher of g-C3N4 via resonance energy transfer (RET) and catalyzing the reduction of O-2, the co-reactant of g-C3N4. Meanwhile, it could promote the ECL of Ru(bpy)(3)(2+) at anode, since the catalytic reduction of O-2 at the cathode increased the faradiac current flowing through the BPE. Based on this signal composite, an ECL off-on phenomenon was observed at the cathode, after Pt-PAMAM-DNAzyme was peeled off by PSA. Conversely, at the anode, an on-off ECL changing was obtained. Therefore, a sensitive ratiometry for PSA detection was achieved with a linear range from 0.10 to 200 ng/mL. Since the two ECL emitters were physically separated, the ratiometric system was relatively simple and neither optical filters nor spectrometer were required. The strategy combining the ECL ratiometry and BPE broadens the applications of BPE-ECL and shows good perspective in clinical application. (C) 2016 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available