Journal
APL MATERIALS
Volume 7, Issue 5, Pages -Publisher
AMER INST PHYSICS
DOI: 10.1063/1.5083810
Keywords
-
Funding
- Fondazione CARIFI Bando
Ask authors/readers for more resources
Inorganic CsPbBr3 perovskite devices have been manufactured and tested as dosimeters under both conventional and Intensity Modulated Radiotherapy (IMRT) X-ray beams. Samples showed a very good linear dependence of the collected charge/current on dose/dose rates in the range of 0.1-5.0 Gy/0.1-4.0 Gy/min of interest for clinical applications. A device sensitivity of about 70 nC/Gy mm(3) compares favorably with other solid-state dosimeters. The first verification of an IMRT dose profile of a prostate cancer treatment, performed by moving the perovskite device on a 10 cm-long profile with a 0.5 mm pitch, showed agreement within 5% with the dose distribution required by the treatment planning system.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available