Journal
WORLD JOURNAL OF GASTROENTEROLOGY
Volume 14, Issue 3, Pages 479-483Publisher
BAISHIDENG PUBLISHING GROUP INC
DOI: 10.3748/wjg.14.479
Keywords
COBAS amplicor test; hepatitis B virus; viral DNA; real-time PCR
Categories
Ask authors/readers for more resources
AIM: To compare the clinical performance of a real-time PCR assay with the COBAS Amplicor Hepatitis B Virus (HBV) Monitor test for quantitation of HBV DNA in serum samples. METHODS: The reference sera of the Chinese National Institute for the Control of Pharmaceutical and Biological Products and the National Center for Clinical Laboratories of China, and 158 clinical serum samples were used in this study. The linearity, accuracy, reproducibility assay time, and costs of the real-time PCR were evaluated and compared with those of the Cobas Amplicor test. RESULTS: The intra-assay and inter-assay variations of the real-time PCR ranged from 0.3% to 3.8% and 1.4% to 8.1%, respectively. The HBV DNA levels measured by the real-time PCR correlated very well with those obtained with the COBAS Amplicor test (r = 0.948). The real-time PCR HBV DNA kit was much cheaper and had a wider dynamic range. CONCLUSION: The real-time PCR assay is an excellent tool for monitoring of HBV DNA levels in patients with chronic hepatitis B. (c) 2008 WJG. All rights reserved.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended