VALIDATION OF DYNAMIC CONTRAST-ENHANCED ULTRASOUND IN RODENT KIDNEYS AS AN ABSOLUTE QUANTITATIVE METHOD FOR MEASURING BLOOD PERFUSION

Journal

ULTRASOUND IN MEDICINE AND BIOLOGY
Volume 37, Issue 6, Pages 900-908

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2011.03.011

Keywords

Destruction-reperfusion; Flash replenishment; Microbubble; Contrast agent; Blood flow; Perfusion

Funding

  1. NIH [HL-02334]
  2. UNC Lineberger Cancer Center

Ask authors/readers for more resources

Contrast-enhanced ultrasound (CEUS) has demonstrated utility in the monitoring of blood flow in tissues, organs and tumors. However, current CEUS methods typically provide only relative image-derived measurements, rather than quantitative values of blood flow in milliliters/minute per gram of tissue. In this study, CEUS derived parameters of blood flow are compared with absolute measurements of blood flow in rodent kidneys. Additionally, the effects of contrast agent infusion rate and transducer orientation on image-derived perfusion measurements are assessed. Both wash-in curve and time-to-refill algorithms are examined. Data illustrate that for all conditions, image-derived flow measurements were well-correlated with transit-time flow probe measurements (R > 0.9). However, we report differences in the sensitivity to flow across different transducer orientations as well as the contrast analysis algorithm utilized. Results also indicate that there exists a range of contrast agent flow rates for which image-derived estimates are consistent. (E-mail: padayton@bme.unc.edu) (C) 2011 World Federation for Ultrasound in Medicine & Biology.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available