MULTIGATE DOPPLER MEASUREMENTS OF ULTRASONIC ATTENUATION AND BLOOD HEMATOCRIT IN HUMAN ARTERIES

Journal

ULTRASOUND IN MEDICINE AND BIOLOGY
Volume 35, Issue 2, Pages 230-236

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2008.08.009

Keywords

Blood; Hematocrit; Doppler; Ultrasonic attenuation

Funding

  1. Polish State Committee for Scientific Research [KBN 5T071303225]

Ask authors/readers for more resources

A clinically applicable method for noninvasive measurement of hematocrit based on 20 MHz multigate Doppler ultrasound was developed. The ultrasound attenuation coefficient in blood is obtained by measuring the power of the signal coming from gates at different depths. A robust averaging method is introduced, which provides stable and repeatable results by using the echo signals from all depths inside the vessel. In vitro measurements have been done on porcine blood with hematocrit ranging from 3.0% to 65.0%. Steady and pulsatile flow conditions have been simulated using a peristaltic pump. The attenuation coefficient indicated the linear relation to hematocrit. The resulting correlation coefficient was R = 0.999 for the continuous blood flow and R = 0.992 for pulsatile flow. In vivo measurements have been performed in the brachial artery in 43 patients with hematocrit in the range of 32.0% to 49.3%. The mean absolute error has been 3.24% with a standard deviation of 3.72%. (E-mail: wsecom@ippt.gov.pl) (c) 2009 World Federation for Ultrasound in Medicine & Biology.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available