Multivariate optimization on flow-injection electrochemical hydride generation atomic absorption spectrometry of cadmium

Journal

TALANTA
Volume 97, Issue -, Pages 229-234

Publisher

ELSEVIER
DOI: 10.1016/j.talanta.2012.04.022

Keywords

Electrochemical hydride generation; Atomic absorption spectrometry; Flow-injection; Experimental design; Cadmium

Ask authors/readers for more resources

A novel electrochemical hydride generation (ECHG) system working in flow-injection (PI) mode was developed for determination of cadmium coupled to an electrically heated quartz tube atomizer (QTA) by atomic absorption spectrometry (AAS). A Plackett-Burman experimental design for screening has been used to evaluate the influence of several variables on the analytical response. Then, the significant parameters such as the concentration of NaCl in catholyte, applied electrolytic current and flow rate of carrier gas have been simultaneously optimized using a central composite design (CCD). Under the optimized conditions, the detection limit (3 sigma(b), n=9) was found to be 0.51 ng mL(-1) Cd and the relative standard deviation (RSD) for nine replicate analyses of 20 ng mL(-1) Cd was 6.5%. The calibration curve was linear in the range of 2-50 ng mL(-1) of Cd. The potential interferences from various ions were also evaluated. The analysis of a reference material showed good agreement with the certified value. The proposed method was successfully applied to the determination of Cd in tap water sample. (C) 2012 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available