RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 23, Issue 9, Pages 1303-1312Publisher
WILEY
DOI: 10.1002/rcm.3983
Keywords
-
Ask authors/readers for more resources
This paper describes an automated workflow for the determination of selected reaction monitoring (SRM) transitions and optimum mass spectrometric (MS) instrument parameters. The approach uses a Nanomate from Advion Biosciences for automated infusion of small amounts of sample in combination with Automaton optimization software from Sciex. The results are stored in the Analyst software Compound Database for automated acquisition method building. Comparisons are presented between the more traditional optimization methods of manual flow injection optimization, Autotune infusion optimization, Automaton flow injection optimization and the Nanomate-Automaton optimization approach. Data is also presented to show that acquisition methods developed on the Sciex model API3000 instrument can be effectively transferred to the Sceix API4000 and API5000 model instruments. Copyright (C) 2009 John Wiley & Sons, Ltd.