4.7 Article

Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom

Journal

RADIOTHERAPY AND ONCOLOGY
Volume 96, Issue 2, Pages 250-253

Publisher

ELSEVIER IRELAND LTD
DOI: 10.1016/j.radonc.2010.03.016

Keywords

Small photon beams; Monte Carlo; Film dosimetry; Stereotactic radiosurgery

Ask authors/readers for more resources

This work reports the use of both GafChromic EBT film immersed in a water phantom and Monte Carlo (MC) simulations for small photon beam stereotactic radiosurgery dosimetry Circularly collimated photon beams with diameters in the 4-20 mm range of a dedicated 6 MV linear accelerator (Novalis (R), Brain-LAB, Germany) were used to perform off-axis ratios, tissue maximum ratios and total scatter factors measurements, and MC simulations GafChromic EBT film data show an excellent agreement with MC results (<2 7%) for all measured quantities (C) 2010 Elsevier Ireland Ltd All rights reserved Radiotherapy and Oncology 96 (2010) 250-253

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available