4.7 Article

Simultaneous MR/PET imaging of the human brain: Feasibility study

Journal

RADIOLOGY
Volume 248, Issue 3, Pages 1028-1035

Publisher

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.2483071927

Keywords

-

Ask authors/readers for more resources

The purpose of this study was to apply a magnetic resonance (MR) imaging-compatible positron emission tomographic (PET) detector technology for simultaneous MR/PET imaging of the human brain and skull base. The PET detector ring consists of lutetium oxyorthosilicate (LSO) scintillation crystals in combination with avalanche photodiodes (APDs) mounted in a clinical 3-T MR imager with use of the birdcage transmit/receive head coil. Following phantom studies, two patients were simultaneously examined by using fluorine 18 fluorodeoxyglucose (FDG) PET and MR imaging and spectroscopy. MR/PET data enabled accurate coregistration of morphologic and multifunctional information. Simultaneous MR/PET imaging is feasible in humans, opening up new possibilities for the emerging field of molecular imaging. (C) RSNA, 2008.