4.0 Article

Application of a strong FSI coupling scheme for the numerical simulation of bileaflet mechanical heart valve dynamics: study of wall shear stress on the valve leaflets

Journal

PROGRESS IN COMPUTATIONAL FLUID DYNAMICS
Volume 12, Issue 2-3, Pages 68-79

Publisher

INDERSCIENCE ENTERPRISES LTD
DOI: 10.1504/PCFD.2012.047450

Keywords

FSI; fluid-structure interaction; algorithm; BMHV; bileaflet mechanical heart valves; biomedical applications

Funding

  1. BOF-grant (Special Research Fund) from Ghent University Association
  2. Research Foundation - Flanders (FWO)

Ask authors/readers for more resources

One of the major challenges in the design of Bileaflet Mechanical Heart Valves (BMHVs) is reduction of the blood damage generated by non-physiological blood flow. Numerical simulations provide relevant insights into the (fluid) dynamics of the BMHV and are used for design optimisation. In this paper, a strong coupling algorithm for the partitioned Fluid-Structure Interaction (FSI) simulation of a BMHV is presented. The convergence of the coupling iterations between the flow solver and the leaflet motion solver is accelerated by using a numerically calculated Jacobian with the derivatives of the pressure and viscous moments acting on the leaflets with respect to the leaflet accelerations. The developed algorithm is used to simulate the dynamics of a 3D BMHV in three different geometries, allowing an analysis of the solution process. Moreover, the leaflet kinematics and the general flow field are discussed, with special focus on the shear stresses on the valve leaflets.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available