4.6 Article

Monte Carlo-based diode design for correction-less small field dosimetry

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 58, Issue 13, Pages 4501-4512

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/58/13/4501

Keywords

-

Funding

  1. Australian Research Council
  2. Queensland University of Technology (QUT)
  3. Wesley Research Institute
  4. Premion [LP110100401]

Ask authors/readers for more resources

Due to their small collecting volume, diodes are commonly used in small field dosimetry. However, the relative sensitivity of a diode increases with decreasing small field size. Conversely, small air gaps have been shown to cause a significant decrease in the sensitivity of a detector as the field size is decreased. Therefore, this study uses Monte Carlo simulations to look at introducing air upstream to diodes such that they measure with a constant sensitivity across all field sizes in small field dosimetry. Varying thicknesses of air were introduced onto the upstream end of two commercial diodes (PTW 60016 photon diode and PTW 60017 electron diode), as well as a theoretical unenclosed silicon chip using field sizes as small as 5 mm x 5 mm. The metric D-w,D-Q/D-Det,(Q) used in this study represents the ratio of the dose to a point of water to the dose to the diode active volume, for a particular field size and location. The optimal thickness of air required to provide a constant sensitivity across all small field sizes was found by plotting D-w,D-Q/D-Det,(Q) as a function of introduced air gap size for various field sizes, and finding the intersection point of these plots. That is, the point at which D-w,D-Q/D-Det,(Q) was constant for all field sizes was found. The optimal thickness of air was calculated to be 3.3, 1.15 and 0.10 mm for the photon diode, electron diode and unenclosed silicon chip, respectively. The variation in these results was due to the different design of each detector. When calculated with the new diode design incorporating the upstream air gap, k(Qclin,Qmsr)(fclin,fmsr) was equal to unity to within statistical uncertainty (0.5%) for all three diodes. Cross-axis profile measurements were also improved with the new detector design. The upstream air gap could be implanted on the commercial diodes via a cap consisting of the air cavity surrounded by water equivalent material. The results for the unclosed silicon chip show that an ideal small field dosimetry diode could be created by using a silicon chip with a small amount of air above it.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available