4.5 Article

Rotationally controlled magneto-hydrodynamic particle handling for bead-based microfluidic assays

Journal

MICROFLUIDICS AND NANOFLUIDICS
Volume 13, Issue 4, Pages 675-681

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s10404-012-0994-7

Keywords

-

Funding

  1. Science Foundation Ireland [10/CE/B1821]

Ask authors/readers for more resources

This work presents a novel magnetic actuation scheme for advanced particle handling on our previously introduced, centrifugal microfluidic platform for array-based analysis of individual cells and beads. The conceptually simple actuation is based on the reciprocating motion of an elastomeric membrane featuring an integrated permanent magnet and a stationary magnet aligned along the orbit of a disc-based chamber. This compression chamber is placed at the downstream end of the particle capture chamber to induce centripetally directed, hydrodynamic lift forces on particles trapped in V-shaped geometrical barriers. Towards high frequencies of rotation, the on-disc magnet ceases to follow the rapidly oscillating magnetic field, so that the magnetic actuator is disabled during the initial, sedimentation-based filling of the trap array. At reduced spin speeds, the residence time of the magnetic actuator is sufficient to displace the magnetic actuator, resulting in a flow through the V-cup array that re-distributes, and eventually fully depletes, the previously trapped beads from the array. The same magnetic deflection scheme is also demonstrated to accelerate mixing, e.g. for upstream sample preparation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available