Shimming with permanent magnets for the x-ray detector in a hybrid x-ray/MR system

Journal

MEDICAL PHYSICS
Volume 35, Issue 9, Pages 3895-3902

Publisher

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2963994

Keywords

XMR hybrid system; magnetic field shimming; permanent magnet

Funding

  1. NIH [RR09784, EB000198]
  2. GE Healthcare
  3. Lucas Foundation
  4. NATIONAL CENTER FOR RESEARCH RESOURCES [P41RR009784] Funding Source: NIH RePORTER
  5. NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING [R01EB000198] Funding Source: NIH RePORTER

Ask authors/readers for more resources

In this x-ray/MR hybrid system an x-ray flat panel detector is placed under the patient cradle, close to the MR volume of interest (VOI), where the magnetic field strength is similar to 0.5 T. Immersed in this strong field, several electronic components inside the detector become magnetized and create an additional magnetic field that is superimposed on the original field of the MR scanner. Even after linear shimming, the field homogeneity of the MR scanner remains disrupted by the detector. The authors characterize the field due to the detector with the field of two magnetic dipoles and further show that two sets of permanent magnets (NdFeB) can withstand the main magnetic field and compensate for the nonlinear components of the additional field. The ideal number of magnets and their locations are calculated based on a field map measured with the detector in place. Experimental results demonstrate great promise for this technique, which may be useful in many settings where devices with magnetic components need to be placed inside or close to an MR scanner. (C) 2008 American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available