4.7 Article

Robotic tissue tracking for beating heart mitral valve surgery

Journal

MEDICAL IMAGE ANALYSIS
Volume 17, Issue 8, Pages 1236-1242

Publisher

ELSEVIER
DOI: 10.1016/j.media.2010.06.007

Keywords

3D ultrasound; Real-time tissue tracking; Motion compensation; Medical robotics; Beating heart surgery

Funding

  1. US National Institutes of Health [NIH R01 HL073647-06]

Ask authors/readers for more resources

The rapid motion of the heart presents a significant challenge to the surgeon during intracardiac beating heart procedures. We present a 3D ultrasound-guided motion compensation system that assists the surgeon by synchronizing instrument motion with the heart. The system utilizes the fact that certain intracardiac structures, like the mitral valve annulus, have trajectories that are largely constrained to translation along one axis. This allows the development of a real-time 3D ultrasound tissue tracker that we integrate with a 1 degree-of-freedom (DOF) actuated surgical instrument and predictive filter to devise a motion tracking system adapted to mitral valve annuloplasty. In vivo experiments demonstrate that the system provides highly accurate tracking (1.0 mm error) with 70% less error than manual tracking attempts. (C) 2010 Elsevier B.V. All rights reserved.