4.4 Article

Validation of a C2-C7 cervical spine finite element model using specimen-specific flexibility data

Journal

MEDICAL ENGINEERING & PHYSICS
Volume 32, Issue 5, Pages 482-489

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2010.03.001

Keywords

Finite element analysis; Cervical spine; Patient-specific modeling

Funding

  1. University of Iowa

Ask authors/readers for more resources

This study presents a specimen-specific C2-C7 cervical spine finite element model that was developed using multiblock meshing techniques. The model was validated using in-house experimental flexibility data obtained from the cadaveric specimen used for mesh development. The C2-C7 specimen was subjected to pure continuous moments up to +/- 1.0 N m in flexion, extension, lateral bending, and axial rotation, and the motions at each level were obtained. Additionally, the specimen was divided into C2-C3, C4-C5, and C6-C7 functional spinal units (FSUs) which were tested in the intact state as well as after sequential removal of the interspinous, ligamentum flavum, and capsular ligaments. The finite element model was initially assigned baseline material properties based on the literature, but was calibrated using the experimental motion data which was obtained in-house, while utlizing the ranges of material property values as reported in the literature. The calibrated model provided good agreement with the nonlinear experimental loading curves, and can be used to further study the response of the cervical spine to various biomechanical investigations. (C) 2010 IPEM. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available