Reduction of the radiofrequency heating of metallic devices using a dual-drive birdcage coil

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 69, Issue 3, Pages 845-852

Publisher

WILEY
DOI: 10.1002/mrm.24316

Keywords

RF safety; transmit arrays; dual-drive birdcage coil; implant heating

Funding

  1. TUBITAK [107E108]

Ask authors/readers for more resources

In this work, it is demonstrated that a dual-drive birdcage coil can be used to reduce the radiofrequency heating of metallic devices during magnetic resonance imaging. By controlling the excitation currents of the two channels of a birdcage coil, the radiofrequency current that is induced near the lead tip could be set to zero. To monitor the current, the image artifacts near the lead tips were measured. The electric field distribution was controlled using a dual-drive birdcage coil. With this method, the lead currents and the lead tip temperatures were reduced substantially [<0.3 degrees C for an applied 4.4 W/kg SAR compared to >4.9 degrees C using quadrature excitation], as demonstrated by phantom and animal experiments. The homogeneity of the flip angle distribution was preserved, as shown by volunteer experiments. The normalized root-mean-square error of the flip angle distribution was less than 10% for all excitations. The average specific absorption rate increased as a trade-off for using different excitation patterns. Magn Reson Med, 2013. (c) 2012 Wiley Periodicals, Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available