Journal
JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY
Volume 22, Issue 9, Pages 1727-1735Publisher
SOC BRASILEIRA QUIMICA
DOI: 10.1590/S0103-50532011000900015
Keywords
anodic stripping voltammetry; carbon paste electrode; 2-aminothiazole-functionalized silica gel; lead and copper determination; water samples
Categories
Funding
- Fundacao de Amparo a Pesquisa do Estado de Minas Gerais (FAPEMIG) [CEX APQ-00291-09]
- Conselho Nacional de Desenvolvimento Cientifico e Tecnologico (CNPq)
- PIBIC/UFU/CNPq
Ask authors/readers for more resources
A solid paraffin-based carbon paste electrode modified with 2-aminothiazole functionalized silica-gel was used for simultaneous quantification of Pb2+ and Cu2+ in water samples by anodic stripping voltammetry. The present method uses short preconcentration time (180 s), which allowed reliable and simultaneous quantification of Pb2+ and Cu2+ in a very fast way. Detection limits of 7.3 and 90 nmol L-1 were obtained for Pb2+ and Cu2+, respectively. These values are below their maximum concentrations allowed in drinking water by Brazilian legislation, indicating that the proposed method is useful to monitoring Pb2+ and Cu2+ in water samples. Recovery studies were performed in four spiked water samples and the results shown recovery percentages close to 100% for all analyzed samples, demonstrating the good accuracy of the proposed method.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available