4.4 Article

Biomechanical evaluation of conventional anulus fibrosus closure methods required for nucleus replacement

Journal

JOURNAL OF NEUROSURGERY-SPINE
Volume 9, Issue 3, Pages 307-313

Publisher

AMER ASSOC NEUROLOGICAL SURGEONS
DOI: 10.3171/SPI/2008/9/9/307

Keywords

anulus closure; anulus sealing; biornechanical testing; intervertebral disc; lumbar spine; nucleus replacement

Funding

  1. Kinetics AG, Esslingen, Germany

Ask authors/readers for more resources

Object. Nucleus replacement implants became increasingly attractive as an alternative to fusion, discectomy, or total disc replacement. The goals of nucleus replacement are to restore disc height and flexibility and to preserve the anatomy. However, implant extrusions have been reported and are the major concern. In this study the authors investigated different conventional surgical methods for anulus closure: suture alone, and fibrin glue and cyanoacrylate glue, alone and with suture. Methods. The in Vitro testing was conducted using 30 lumbar spinal segments obtained from calves. In each specimen, an incision was made; the nucleus was removed and subsequently replaced by a collagen matrix. The incisions were treated with anulus closure methods in 5 groups of animals. Flexibility was assessed in a spine tester. Subsequently, specimens were exposed to cyclic fatigue loading by using a hydraulic loading frame. Specimens were excentrically loaded in sine waveform up to a maximum of 100,000 cycles with 4-24 Nm at 5 Hz while being rotated at 360 degrees/minute. Results. Removal of the nucleus caused a significant loss of stability. The segmental stability could be restored after the implantation. Fatigue testing indicated that suturing was able to sustain 3400 cycles. Fibrin glue failed earlier than cyanoacrylate glue. Both combinations (suture with glue) provided longer stability to the anulus closure. Conclusions. The results suggested that closing the anulus incision with suture or fibrin glue alone might not be appropriate. The authors found that the best method was cyanoacrylate glue with suture. Although this method provided the longest duration of closure, it could not sustain the maximum number of fatigue cycles. Conventional methods could improve the outcome compared with using no closure. Nonetheless, the authors' findings highlight the demand for an appropriate anulus reconstruction method or device with good long-term reliability.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available