4.1 Article

VALIDATION THIN LAYER CHROMATOGRAPHY FOR THE DETERMINATION OF NAPROXEN IN TABLETS AND COMPARISON WITH A PHARMACOPEIL METHOD

Journal

JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES
Volume 34, Issue 10-11, Pages 829-847

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/10826076.2011.571152

Keywords

densitometry; naproxen; pharmaceutical analysis; quantitative analysis; spectrodensitometry; spectrophotometry; TLC

Funding

  1. Medical University of Silesia [KNW-1-001/10]

Ask authors/readers for more resources

Adsorption thin layer chromatography with densitometry has been established for the identification and the quantification of naproxen in tablets. UV densitometry was performed in absorbance mode at 230nm. The presented method was validated by specificity, range, linearity, accuracy, precision, detection limit, quantitative limit, and robustness. The TLC-densitometric method was also compared with a pharmacopeial UV-spectrophotometric method for the assay of naproxen, and the results confirmed statistically that the TLC-densitometric method can be used as a substitute method.