Evaluation of the radial artery applanation tonometry technology for continuous noninvasive blood pressure monitoring compared with central aortic blood pressure measurements in patients with multiple organ dysfunction syndrome

Journal

JOURNAL OF CRITICAL CARE
Volume 28, Issue 6, Pages 908-912

Publisher

W B SAUNDERS CO-ELSEVIER INC
DOI: 10.1016/j.jcrc.2013.06.012

Keywords

Mean arterial pressure; Systolic blood pressure; Diastolic blood pressure; Hemodynamic monitoring; T-Line

Funding

  1. Tensys Medical Inc (San Diego, Calif)

Ask authors/readers for more resources

Purpose: We compared blood pressure (BP) measurements obtained using radial artery applanation tonometry with invasive BP measurements using a catheter placed in the abdominal aorta through the femoral artery in patients with multiple organ dysfunction syndrome (MODS). Materials and Methods: In 23 intensive care unit patients with MODS, we simultaneously assessed BP values for 15 minutes per patient using radial artery applanation tonometry (T-Line TL-200pro device; Tensys Medical Inc, San Diego, Calif) and the arterial catheter (standard-criterion technique). A total of 2879 averaged 10-beat epochs were compared using Bland-Altman plots. Results: The mean difference +/- SD (with corresponding 95% limits of agreement) between radial artery applanation tonometry-derived BP and invasively assessed BP was + 1.0 +/- 5.5 mm Hg (-9.9 to + 11.8 mm Hg) for mean arterial pressure, -3.3 +/- 11.2 mm Hg (-25.3 to + 18.6 mm Hg) for systolic arterial pressure, and + 4.9 +/- 7.0 mm Hg (-8.8 to + 18.6 mm Hg) for diastolic arterial pressure, respectively. Conclusions: In intensive care unit patients with MODS, mean arterial pressure and diastolic arterial pressure can be determined accurately and precisely using radial artery applanation tonometry compared with central aortic values obtained using a catheter placed in the abdominal aorta through the femoral artery. Although systolic arterial pressure could also be derived accurately, wider 95% limits of agreement suggest lower precision for determination of systolic arterial pressure. (C) 2013 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available