Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT film

Journal

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS
Volume 9, Issue 3, Pages 90-98

Publisher

WILEY
DOI: 10.1120/jacmp.v9i3.2802

Keywords

mMLC; MLC transmission; MLC penumbra; GafChromic; dosimetry; IMRT

Funding

  1. Conacyt [U46761-F]

Ask authors/readers for more resources

Micro-multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation system and requires the use of detectors with high spatial resolution, high sensitivity and practically no energy dependence. In this work the use of GafChromic EBT radiochromic film to measure the basic dosimetric properties of the m3-mMLC (BrainLAB, Germany) micro-multileaf collimator system integrated to a 6 MV linear accelerator, is reported. Results show that average values of transmission and leakage radiation are 0.93 +/- 0.05% and 1.08 +/- 0.08%, respectively. The 80-20% beam penumbra were found to be 2.26 +/- 0.11 mm along the leaf side (perpendicular to leaf motion) and 2.31 +/- 0.11 mm along the leaf end ( parallel to leaf motion) using square field sizes ranging from 9.1 to 1.8 cm. These measurements are in agreement with values reported in the literature for the same type of mMLC using different radiation detectors.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available