INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY
Volume 4, Issue 2, Pages 180-191Publisher
WILEY
DOI: 10.1002/rcs.184
Keywords
3DUS; medical robots; IGT; CIS; liver; ablation; RFA
Categories
Funding
- NCI NIH HHS [R41 CA103468, R44 CA134169, 5-1R41CA103468-01] Funding Source: Medline
Ask authors/readers for more resources
Background Clinical use of image-guided needle placement robots has lagged behind laboratory-demonstrated robotic capability. Bridging this gap requires reliable and easy-to-use robotic systems. Methods Our system for image-guided needle placement requires only simple, low-cost components and minimal, entirely off-line calibration. It rapidly aligns needles to planned entry paths using 3D ultrasound (US) reconstructed from freehand 2D scans. We compare system accuracy against clinical standard manual needle placement. Results The US-guided robotic system is significantly more accurate than single manual insertions. When several manual withdrawals and reinsertions are allowed, accuracy becomes equivalent. In ex vivo experiments, robotic repeatability was 1.56 mm, compared to 3.19 and 4.63 mm for two sets of manual insertions. In an in vivo experiment with heartbeat and respiratory effects, robotic system accuracy was 5.5 mm. Conclusions A 3D US-guided robot can eliminate error bias and reduce invasiveness (the number of insertions required) compared to manual needle insertion. Remaining future challenges include target motion compensation. Copyright (c) 2008 John Wiley & Sons, Ltd.