4.2 Article

Third-body abrasive wear of tibial polyethylene inserts combined with metallic and ceramic femoral components in a knee simulator study

Journal

INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS
Volume 36, Issue 1, Pages 47-55

Publisher

WICHTIG EDITORE
DOI: 10.5301/ijao.5000189

Keywords

Total knee arthroplasty; Tribology; Knee simulator; Third-body wear; Ceramics

Ask authors/readers for more resources

Aim: Total knee arthroplasties have reached a high grade of quality and safety, but most often fail because of aseptic implant loosening caused by polyethylene wear debris. Wear is generated at the articulating surfaces, e. g. caused by third-body particles. The objective of this experimental study was to determine the wear of tibial polyethylene inserts combined with metallic and ceramic femoral components under third-body wear conditions initiated by bone cement particles. Methods and Materials: Wear testing using a cemented unconstrained bicondylar knee endoprosthesis (Multigen Plus CR knee system) was performed in a knee wear simulator. Tibial polyethylene inserts were combined with the identical femoral component design, but made of two different materials (cobalt-chromium and ceramic). Bone cement debris including zirconium oxide particles was added every 500,000 cycles between the articulating surfaces. After 5 million load cycles, the amount of wear was determined gravimetrically and compared with results from standard wear test conditions. The surfaces of tibial inserts were also analyzed. Results: The average gravimetrical wear of the tibial polyethylene inserts in combination with cobalt-chromium and ceramic femoral components under third-body wear conditions amounted to 31.88 +/- 4.53 mg and 13.06 +/- 1.88 mg after 5 million cycles, respectively, and was higher than under standard wear test conditions in both cases. Conclusions: The wear simulator test demonstrates that wear of polyethylene inserts under third-body wear conditions, in combination with ceramic femoral components, was significantly lower than with metallic femoral components.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available