4.0 Article

Hb Belleville [10(A7)AlaThr] Affects The Determination of Hb A1c by Routine Cation Exchange High Performance Liquid Chromatography

Journal

HEMOGLOBIN
Volume 33, Issue 1, Pages 45-50

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/03630260802625915

Keywords

Hb A1c; Hemoglobin (Hb) variant; Electrospray ionization mass spectrometry (ESI-MS)

Funding

  1. Wellcome Trust Funding Source: Medline

Ask authors/readers for more resources

When analyzed by routine cation exchange high performance liquid chromatography (HPLC), the Hb A1c peak of a Caucasian diabetic patient from Yorkshire, UK, appeared to be an incompletely resolved doublet. One component (5.5%) eluted at the normal time for Hb A1c, whereas the other component (6.6%) eluted slightly later. The HPLC trace was otherwise normal. Analysis of the diabetic patient's blood and a tryptic digest thereof by electrospray ionization mass spectrometry (ESI-MS) identified the Hb Belleville trait. To relate Hb A1c determined by HPLC to and chain glycation determined by ESI-MS, several normal blood samples (5-14% Hb A1c) were analyzed by both techniques. The Hb A1c levels derived from the and chain glycation levels of the diabetic patient (12.9 and 12.4%, respectively) agreed with the sum of the two peaks (12.1%) in the HPLC trace. Similarly, Hb Belleville was detected and identified in another Caucasian, also from Yorkshire, with normal Hb A1c.