Influence and reliability of lower-limb arterial occlusion pressure at different body positions

期刊

PEERJ
卷 6, 期 -, 页码 -

出版社

PEERJ INC
DOI: 10.7717/peerj.4697

关键词

Blood flow restriction exercise; Ischemic preconditioning; Occlusion; Limb occlusion pressure; Kaatsu

资金

  1. Delfi Medical

向作者/读者索取更多资源

Background: Total arterial occlusive pressure (AOP) is used to prescribe pressures for surgery, blood flow restriction exercise (BFRE) and ischemic preconditioning (IPC). AOP is often measured in a supine position; however, the influence of body position on AOP measurement is unknown and may influence level of occlusion in different positions during BFR and IPC. The aim of this study was therefore to investigate the influence of body position on AOP. Methods: Fifty healthy individuals (age = 29 +/- 6 y) underwent AOP measurements on the dominant lower-limb in supine, seated and standing positions in a randomised order. AOP was measured automatically using the Delfi Personalised Tourniquet System device, with each measurement separated by 5 min of rest. Results: Arterial occlusive pressure was significantly lower in the supine position compared to the seated position (187.00 +/- 32.5 vs 204.00 +/- 28.5 mmHg, p < 0.001) and standing position (187.00 +/- 32.5 vs 241.50 +/- 49.3 mmHg, p < 0.001). AOP was significantly higher in the standing position compared to the seated position (241.50 +/- 49.3 vs 204.00 +/- 28.5 mmHg, p < 0.001). Discussion: Arterial occlusive pressure measurement is body position dependent, thus for accurate prescription of occlusion pressure during surgery, BFR and IPC, AOP should be measured in the position intended for subsequent application of occlusion.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据