4.3 Review

Instrumented hip joint replacements, femoral replacements and femoral fracture stabilizers

期刊

EXPERT REVIEW OF MEDICAL DEVICES
卷 11, 期 6, 页码 617-635

出版社

EXPERT REVIEWS
DOI: 10.1586/17434440.2014.946695

关键词

arthroplasty; energy harvesting; femoral fixation; femoral replacement; hip joint replacement; instrumented implant; medical device; orthopedics; telemetry; therapeutic actuator

资金

  1. Portuguese Foundation for Science and Technology (FCT) [PTDC/EME-PME/105465/2008, SFRH/BD/78414/2011]
  2. Fundação para a Ciência e a Tecnologia [SFRH/BD/78414/2011, PTDC/EME-PME/105465/2008] Funding Source: FCT

向作者/读者索取更多资源

This paper reviews instrumented hip joint replacements, instrumented femoral replacements and instrumented femoral fracture stabilizers. Examination of the evolution of such implants was carried out, including the detailed analysis of 16 architectures, designed by 8 research teams and implanted in 32 patients. Their power supply, measurement, communication, processing and actuation systems were reviewed, as were the tests carried out to evaluate their performance and safety. These instrumented implants were only designed to measure biomechanical and thermodynamic quantities in vivo, in order to use such data to conduct research projects and optimize rehabilitation processes. The most promising trend is to minimize aseptic loosening and/or infection following hip or femoral replacements or femoral stabilization procedures by using therapeutic actuators inside instrumented implants to apply controlled stimuli in the bone-implant interface.