Design and Performance Evaluation of a Linear Ion Trap Mass Analyzer Featuring Half Round Rod Electrodes

期刊

出版社

SPRINGER
DOI: 10.1007/s13361-015-1085-2

关键词

Linear ion trap; Radial ejection; Mass analyzer; Round rod electrode; Mass Resolution

资金

  1. Natural Science Foundation of Jiangsu Province of China [BK20140340]
  2. Natural Science Foundation for Youth of Soochow University [SDY2013A11]

向作者/读者索取更多资源

A novel linear ion trap mass analyzer featuring half round rod electrodes (HreLIT) has been built. It is mainly composed of two pairs of stainless steel electrodes which have a cross-section of half round rod and a pair of end electrodes. The HreLIT has a simple structure and so it could be assembled by hand with relatively high mechanical accuracy. The external dimension of HreLIT is 50 mm x 29.5 mm x 28 mm (length x width x height) and its internal volume is about 3.8 cm(3). A home-made HreLIT mass spectrometer with three-stage vacuum system was built and the performance of HreLIT was characterized using reserpine solution and PPG standard solution. When the scan rate was 254 u/s, mass peak with FWHM of 0.14 u was achieved for ions with m/z 609, which corresponds to a mass resolution of 4350. The HreLIT was also operated at a low q value of 0.28 to extend its mass range. The experiment result showed a mass range of over 2800 u and the amplitude of radio frequency (rf) signal was only 1560 V (0-p). Three-stage tandem mass spectrometry was successfully performed in the HreLIT, and the collision-induced dissociation (CID) efficiencies of MS2 (CID of ions with m/z 609) and MS3 (CID of ions with m/z 448) were 78% and 59%, respectively.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据