Image-guided spinal injection procedures in open high-field MRI with vertical field orientation: feasibility and technical features

期刊

EUROPEAN RADIOLOGY
卷 20, 期 2, 页码 395-403

出版社

SPRINGER
DOI: 10.1007/s00330-009-1567-3

关键词

Interventional MRI; Open high field; Spinal injection therapy; Artefacts; Low back pain

向作者/读者索取更多资源

We prospectively evaluated the feasibility and technical features of MR-guided lumbosacral injection procedures in open high-field MRI at 1.0 T. In a CuSO(4)center dot 5H(2)O phantom and five human cadaveric spines, fluoroscopy sequences (proton-density-weighted turbo spin-echo (PDw TSE), T1w TSE, T2w TSE; balanced steady-state free precession (bSSFP), T1w gradient echo (GE), T2w GE) were evaluated using two MRI-compatible 20-G Chiba-type needles. Artefacts were analysed by varying needle orientation to B(0), frequency-encoding direction and slice orientation. Image quality was described using the contrast-to-noise ratio (CNR). Subsequently, a total of 183 MR-guided nerve root (107), facet (53) and sacroiliac joint (23) injections were performed in 53 patients. In vitro, PDw TSE sequence yielded the best needle-tissue contrasts (CNR = 45, 18, 15, 9, and 8 for needle vs. fat, muscle, root, bone and sclerosis, respectively) and optimal artefact sizes (width and tip shift less than 5 mm). In vivo, PDw TSE sequence was sufficient in all cases. The acquisition time of 2 s facilitated near-real-time MRI guidance. Drug delivery was technically successful in 100% (107/107), 87% (46/53) and 87% (20/23) of nerve root, facet and sacroiliac joint injections, respectively. No major complications occurred. The mean procedure time was 29 min (range 19-67 min). MR-guided spinal injections in open high-field MRI are feasible and accurate using fast TSE sequence designs.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据