Quadrupolar ion excitation for radiofrequency-only mass filter operation

期刊

EUROPEAN JOURNAL OF MASS SPECTROMETRY
卷 20, 期 3, 页码 207-214

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1255/ejms.1278

关键词

linear quadrupole; mass selective axial ejection; quadrupolar excitation; mass resolution; excitation time; excitation amplitude

资金

  1. Natural Sciences and Engineering Research Council of Canada
  2. AB Sciex through an Industrial Research Chair

向作者/读者索取更多资源

Trajectory calculations are used to model a mass filter based on the radiofrequency (rf)-only operation of a linear quadrupole with resonant quadrupole excitation of ions (resonant excitation applied with the same spatial electric field as the main quadrupole rf field). Ions are not trapped, but pass continuously through the quadrupole. Excited ions gain axial kinetic energy in the fringe field at the quadrupole exit, overcome a stopping potential and are transmitted to an external detector. No quadrupole direct current is required, unlike conventional operation at the tip of the first stability diagram. Quadrupole excitation can be applied with amplitude or frequency modulation of the main rf voltage, or with an auxiliary excitation voltage. All three methods give the same mass resolution. The mass resolution, R, is given by R approximate to 0.5q(d beta/dq)n where q is a Mathieu parameter, beta(q), determines the frequency of ion oscillation and n is the number of cycles of the rf field experienced by an ion, determined by the flight time through the quadrupole. A disadvantage of this mode of operation is that the flight times of the ions and the excitation amplitudes or modulation depths need to be synchronized.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据