Tachycardia detection performance of implantable loop recorders: results from a large real-life patient cohort and patients with induced ventricular arrhythmias

期刊

EUROPACE
卷 15, 期 8, 页码 1215-1222

出版社

OXFORD UNIV PRESS
DOI: 10.1093/europace/eut036

关键词

Implantable loop recorder; ECG monitoring; Ventricular tachycardia; Ventricular fibrillation; Arrhythmia detection; Algorithms

资金

  1. Medtronic, Inc.

向作者/读者索取更多资源

Implantable loop recorders (ILRs) are valuable for diagnosing arrhythmias. We evaluated tachycardia detection performance of the Medtronic Reveal ILR with FullView Software. The rate of occurrence of tachycardia detection [supraventricular tachycardia, ventricular tachycardia (VT), and ventricular fibrillation (VF)] and the percentage of appropriately detected tachycardias were determined from all 2190 ILR patients that transmitted to CareLink over a 4-month period (total follow-up 135.6 patient-years). All 1909 tachycardia episodes were reviewed. Episodes with actual heart rate above the programmed tachycardia detection rate were classified as appropriate. Sensitivity to detect true ventricular arrhythmias was assessed in another group of 215 patients undergoing implantable cardioverter defibrillator (ICD) implant testing. Skin electrodes represented ILR electrodes. Induced VF (404 episodes) and VT (93 episodes) were processed by an emulation of FullView Software. Generalized estimation equation analysis adjusted for multiple episodes per patient. In the CareLink cohort, 68.7 (63.9 adjusted) of detected episodes had tachycardia above the detection rate. Of 1642 episodes detected in the VT zone (12.1 episodes/patient-year), 78.8 (79.0 adjusted) had tachycardia above the detection rate. Of 267 episodes detected in the fast VT zone (1.9 episodes/patient-year), 6.7 (9.4 adjusted) had tachycardia above the detection rate. Twelve true VT/VF episodes were observed in 10 patients. In the ICD patient cohort, 95.9 (96.5 adjusted) of induced VT/VF segments were correctly detected at nominal rate cutoffs. When VT detection was set to 130 b.p.m. (to include the slowest VT), 99.0 (99.3 adjusted) were correctly detected. The majority (63.9) of detected tachycardias contained true tachycardia. Sensitivity to detect induced VT/VF was 99.3.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据