期刊
ELECTROANALYSIS
卷 22, 期 15, 页码 1765-1771出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200900644
关键词
PGMA; Ferrocene; Glucose oxidase; Biosensors
资金
- T. R Prime Ministry State Planning Organization
- Faith University Research Support Office [P50090801-1]
A novel glucose biosensor was constructed via direct covalent attachment of glucose oxidase onto epoxy group containing polymeric election transfer mediator. Poly(glycidyl methacrylate-co-vinylferrocene) A copolymer of glycidyl methacrylate (GMA) and vinylferrocene (VFc) with different molar rations has been prepared by free radical copolymerization These copolymers have been utilized as polymeric mediators for amperometric glucose sensing The catalytic electrochemistry of the enzyme electrode with the copolymer was investigated Copolymei acts as an electron transfer mediator between the redox center of Glucose oxidase (GOx) and the electrode The stability. reusability. pH and temperature response of the biosensor as well as its kinetic parameter have also been studied
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据