Effect of object location on the density measurement and Hounsfield conversion in a NewTom 3G cone beam computed tomography unit

期刊

DENTOMAXILLOFACIAL RADIOLOGY
卷 37, 期 6, 页码 305-308

出版社

BRITISH INST RADIOLOGY
DOI: 10.1259/dmfr/65993482

关键词

Hounsfield value; computed tomography; bone density

向作者/读者索取更多资源

Objectives: The purpose of this study was to determine the effect of an object's location in a cone beam CT imaging chamber (CBCT-NewTom 3G) on its apparent density and to develop a linear conversion coefficient for Hounsfield units (HU) to material density (g cm(-3)) for the NewTom 3G Scanner. Methods: Three cylindrical models of materials with different densities were constructed and scanned at five different locations in a NewTom 3G Volume Scanner. The average HU value for each model at each location was obtained using two different types of software. Next, five cylinders of different known densities were scanned at the exact centre of a NewTom 3G Scanner. The collected data were analysed using the same two types of software to determine a standard linear relationship between density and HU for each type of software. Results: There is no statistical significance of location of an object within the CBCT scanner on determination of its density. A linear relationship between the density of an object and the HU Of a scan was rho = 0.001(HU)+1.19 with an R2 value of 0.893 (where density, rho, is measured in cm(-3)). This equation is to be used on a range between 1.42 g cm(-3) and 0.4456) g cm(-3). Conclusions: A linear relationship can be used to determine the density of materials (in the density range of bone) from the HU values of a CBCT scan. This relationship is not affected by the object's location within the scanner itself.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据